News

 
  1. RTI Electronics AB is Looking for New Employees

    26 Jan 12 Would you like to be part of developing the medical technology products and markets of the future? RTI Electronics is growing and is now looking for new co-workers who want to be part of shaping the future in an exciting industry.

  2. RTI Electronics, Inc. has Moved

    02 Jan 12 RTI Electronics, Inc. has moved to larger location with lots of space to grow. Starting Tuesday, January 3, 2012 you can find RTI Electronics, Inc. at:

    33 Jacksonville Road
    Building 1
    Towaco, NJ 07082

    RTI Inc's phone numbers and contact information are the same for your convenience.

  3. Merry Christmas 2011 from RTI

    All the Best for 2012

    22 Dec 11 We wish to thank our customers and users for a fruitful 2011, wish you a really nice Christmas, and the best for 2012. Monday 26 December RTI Headquarters is closed.

  4. Ocean-Central-nov-2011webb-2.jpg

    Awaited RTI’s Ocean Central is Here

    07 Dec 11 Ocean Central makes it possible to tie all Ocean users together via network and internet. With Ocean Central you can save all measured data in one place and easily distribute templates to all users.

Show all news

RFM

RFM

The X-ray tube in focus

The RFM (Real-time FocusMeter) is a tool for determining the X-ray tube focal spot in a quick and easy way. No film is needed and the method is simple to apply. The greatest advantage of using this technique is the speed - in less than one minute a complete examination is carried out which gives you both the size and shape for different generator settings. The result of a measurement is available "at once" and can be examined and displayed on a computer screen and the method makes it possible to measure, for instance, the focal spot width and length for different angles from the radiation beam axis. With this method it is also possible to measure focal spot down to 100 µm.

Methods and Automatic Alignment

The RFM is based on the slit method according to the IEC 336 recommendation. The 10 µm wide slit makes it possible to measure focal spot sizes down to 0.10 mm, according to IEC 60336.

The RFM can be automatically aligned. With a specially designed alignment pattern, and four CCD arrays, the RFM determines its position in the beam. It also calculates the enlargement factor and the distance from the focal spot to the detector.


Main features
  • Fast and accurate
  • Automatic alignment
  • Focal spot size
  • Shape (Line Spread Function)
  • Modulation Transfer Function, MTF
  • Root Mean Square (RMS) calculations
  • Microsoft Access based database for saving data
  • Export function for LSF data
  • Only one exposure per axis
  • High sensitivity
  • Mammo focal spot sizes down to 0.1 (IEC 60336)
  • No film needed
  • Real-time measurements
  • Windows software
Specifications general
Supporting standards: IEC 60336

DIN 6823

NEMA XR5
Tube focal spot types: Radiographic

CT

Mammographic

Dental
Exposure time: minimum 10 ms
kV range: 22-150 kV
Charge range, tube: 10-100 mAs (at 70 cm)
Enlargement factor: 0.8-5 (automatically calculated)
Measurement range: 0.1-5mm
Resolution: 25/E µm (E=enlargement calculated)
MTF magn. Factor: 1 < M =< 10
Measured values: focus IEC or DIN or NEMA

focus RTI (length and width)

MTF at -3 dB, M=2

RMS (Root Mean Square)

Zoom mode with user controlled cursors for measurements on LFS and MTF graphs.
Alignment system:  Automatic
Trigger system
Trigger modes Automatic  (with external trig detector)

Auto start, adjustable delay & window

Manual start, adjustable window
Delay 0-50 000 ms
Window 10-50 000 ms
Accessories

R100 Trig detector To start and stop measurements automatically.
RFM Holder pro Holder for the Real-time FocusMeter.

The RFM Pro consists of a modified Real-time FocusMeter (RFM) and the RFM Holder Pro. In the RFM Pro the motors for xy-movement are removed. That makes the positioning of the detector more accurate according to the tube housing. The RFM Pro is intended for use with the RFM always mounted in the holder. The RFM Holder Pro is used by world leading X-ray tube manufacturers. It expands the use of the RFM when a higher degree of accuracy and repeatability of the positioning is required for example when doing absolute measurements in an X-ray tube production line.

 
 

Please enter your user name and password below.

If you are the proud owner of a Barracuda, Piranha, or Nova you just have to enter your serial number as the user name and leave the password box empty.