The different Piranha models are designed as a truly self-contained, all-in-one, multi-function meter and to be smart in different areas, so you can purchase the model that meets your exact application needs. Thanks to the variety of tailor made models and that you have the possibility to upgrade your Piranha – when your needs are growing – your investment is protected.
The Piranha comes ready-to-use. Everything you need is included. Standard detectors are built-in. Communication is made easy via Bluetooth or USB. Just connect the Piranha to your laptop and you have a complete QA-system.
kV, time, dose, dose rate, HVL, and total filtration. The Piranha will also measure CTDI and give you CT dose profiles. With Pulsed Fluoroscopy you add pulse rate and dose/pulse. It measures on Cine and Dental, both Panoramic and Intra-Oral, as well as on traditional and slot scanning Mammography.
The Piranha has an extremely sensitive detector, capable of measuring on low-output mini C-arms and Mammography. The Auto-Compensation pioneered by RTI, avail-able on all our existing multi-function meters, corrects for any added filtration. This ensures confidence in the accuracy and reproducibility of measurements without compromise.
Piranha has a unique feature that makes it possible to check the position of the detector before a measurement is performed. When the Piranha is calibrated, a special filter (built into Piranha) is used to store a reference value with correctly irradiated detectors. By using the "Position Check" function the position of the Piranha can be verified by comparison against the reference value.
Note that this method is superior to other positioning methods used by other devices. When positioning the Piranha, the actual detectors that measure kVp and dose are used to verify the position.
The Piranha models are divided into the following main groups:
The size of the detector is very important when doing measurements on panoramic (OPG) systems. Panoramic systems has a very narrow beam and the detector must be small to be completely irradiated. The sensitive detector area on the Piranha, including kVp and dose detector, is only 3 x 21.1 mm (0.11" x 0.83").
Light green triangle: 3 mm Al + 2.0 mm Cu
Dark green square: 3 mm Al + 1.5 mm Cu
| Type | R100B |
| Sensitivity | 55 µC/Gy |
| Size | 20 x 45 x 7.4 mm, 0.79" x 1.8" x 0.29" |
| Weight 85 g (3 oz) | |
| Active detector area | 10 x 10 mm, 0.39" x 0.39" |
| Cable length | 2.0 m (6.6 ft) |
| Connector | Piranha type with automatic detector identification |
| Dose range | 0.1 nGy - >100 kGy |
| 12 nR - >11000 kR | |
| Dose rate range | 4 nGy/s - 76 mGy/s |
| 1.6 mR/h - 31 kR/h | |
| Dose per pulse | 1 nGy/pulse - 3 kGy/pulse |
| 12 nR/pulse - 340 kR/pulse | |
| Backscatter |
None |
Range
Inaccuracy
35 - 155 kVp (Rad/Flu)
±1.5 %
35 - 105 kVp (Dent) ±1.5 %
±1.5 %
75 - 150 kVp (CT) ±1.5 %
±1.5 %
0.1 ms - 2000 s
±1 % or ±0.5 ms
1 - 65535 pulses
±1 pulse
30 nGy - 1000 Gy
±5 %
4 µR - 100 kR
±5 %
15 nGy/s - 450 mGy/s
±5 % or ±7 nGy/s
1.7 µR/s - 50 R/s
±5 % or ±0.8 µR/s
0.1 mR/min - 3000 R/min
±5 % or ±0.05 mR/min
1.5 - 38 mm Al Total Filtr.
±10 % or ±0.3 mm
1.2 - 14 mm Al HVL
±10 % or ±0.2 mm
| Range Inaccuracy |
| 18 - 49 kVp (Mo/Mo) ±0.7 kV |
| 0.1 ms - 2000 s ±1 % or ±0.5 ms |
| 1 - 65535 pulses ±1 pulse |
| 70 nGy - 2000 Gy ±5 % |
| 8 µR - 200 kR ±5 % |
| 15 nGy/s - 750 mGy/s ±5 % or ±0.04 µGy/s |
| 17 µR/s - 86 R/s ±5 % or ±4 µR/s |
| 1 mR/min - 5100 R/min ±5 % or ±0.3 mR/min |
Mo/Mo, Mo/Rh, Mo/Al, Rh/Rh, Rh/Al, W/Al, W/Rh, W/Ag