News

 
  1. RTI Electronics AB is Looking for New Employees

    26 Jan 12 Would you like to be part of developing the medical technology products and markets of the future? RTI Electronics is growing and is now looking for new co-workers who want to be part of shaping the future in an exciting industry.

  2. RTI Electronics, Inc. has Moved

    02 Jan 12 RTI Electronics, Inc. has moved to larger location with lots of space to grow. Starting Tuesday, January 3, 2012 you can find RTI Electronics, Inc. at:

    33 Jacksonville Road
    Building 1
    Towaco, NJ 07082

    RTI Inc's phone numbers and contact information are the same for your convenience.

  3. All the Best for 2012

    22 Dec 11 We wish to thank our customers and users for a fruitful 2011, wish you a really nice Christmas, and the best for 2012. Monday 26 December RTI Headquarters is closed.

  4. Awaited RTI’s Ocean Central is Here

    07 Dec 11 Ocean Central makes it possible to tie all Ocean users together via network and internet. With Ocean Central you can save all measured data in one place and easily distribute templates to all users.

Show all news

News

2010-09-06

Receive Higher Accuracy when Measuring CTDIw

RTI strongly recommend the 1 point method when measuring CTDIw. Using the 1 point method only one exposure in the centre hole is performed for each CTDIw estimation. This is possible by using know relations in the dose distribution between the centre hole and the 4 periphery holes. This relation is defined by the "k-factor".

How accurate is the k-factor?

The k-factors have been extracted from statistics during two or three years  from measurements of the CTDIw and CTDI100 measured with ion chambers on different scanners.

The variation between different scanners and bowtie filters are less than ±5%. This difference gives less distribution to the expanded uncertainty than influences by measuring the 5 periphery holes individually. Factors that add inaccuracy when measuring in the periphery holes are energy dependencies, low signal and high dynamic variation in signal during the tube rotation. When using the CT dose profiler in the periphery holes during a scan the pitch has to be less than 0,5 to avoid loss of information for the dose estimation.

Can I check the k-factor myself?

Yes; You can easily extract you k-factor for each of your CT scanners by using previous measurements with ion chambers.

RTI strongly recommends the 1 point method to be used for CTDIw measurements with the CT Dose Profiler. The method saves time and gives a higher accuracy than the 5 point method, either using a pencil ion chamber or the CT Dose Profiler.

 