News

 
  1. Receive Higher Accuracy when Measuring CTDIw

    06 Sep 10 RTI strongly recommends the 1-point method to be used for CTDIw measurements with the CT Dose Profiler. The method saves time and gives a higher accuracy than the 5 point method, either using a pencil ion chamber or the CT Dose Profiler.

  2. Decotron – New Distributor in Norway

    05 Sep 10 We welcome our new Norwegian distributor, Decotron, to our RTI Team.

  3. Meet Us at RPM in Bulgaria

    01 Sep 10 RTI will be at RPM in Bulgaria together with our distributor – Ecomed Services – to show you the best from RTI Electronics, our X-ray multimeters Piranha and Barracuda that togehter with our powerful software gives you an extraordinary X-ray QA solution.

  4. assets/News/Juli-2010/_resampled/SetWidth190-South-2.jpg

    RTI Electronics Strengthening its Position in South America

    23 Jul 10 We are now pleased to present three new distributors in South America.

  5. assets/News/Juli-2010/_resampled/SetWidth190-Picture-T20-034.jpg

    Important Solid State Detector Accessories by RTI Electronics

    23 Jul 10 If you own an RTI Piranha or Barracuda multi-meter, you may want to consider adding two special solid state detectors specially developed by RTI Electronics.

  6. EDGE July 2010

    21 Jul 10 Newsletter from RTI Electronics, July issue 2010.

Show all news
Detector Extender

Barracuda Dose Detector R100B

Barracuda Dose Detector R100B

Especially Designed for Low Dose Rate Measurements

The R100B is a dose detector for the Barracuda. It is specially designed for low dose rate measurements. The detector is small to minimize interference with the X-ray systems automatic exposure control. R100B is a solid state detector with a metal housing which makes it very durable. It does not need correction for temperature or pressure and need no bias voltage. The R100B works with all type of electrometer modules available for the Barracuda. Best performance is obtained with the EMM-BiasW module.


Fluoroscopy

The R100B dose detector is designed for low dose rate measurements when measuring the input dose to the image intensifier. It can be used both for continuous and pulsed fluroroscopy. The R100B is small and has a fast response which makes it ideal for pulsed fluoroscopy. It can detect the individual pulses, determine pulse rate and show waveforms even at the highest pulse rates used on modern fluoroscopy systems. This is not possible with ionization chambers since their response is too slow to resolve individual pulses.

Scattered and Leakage Radiation

The R100B used with the EMM-BiasW electrometer module can detect extremely low dose rate levels. A lower range of 1 nGy/s (0.4 mR/h) makes it possible to use it also for measurement of scattered and leakage radiation.

Light green triangle: 3 mm Al + 2.0 mm Cu
Dark green square: 3 mm Al + 1.5 mm Cu

Specifications (Dose Detector R100 B used with EMM Bias W)
Sensitivity  55 µC/Gy
Size 20 x 45 x 7.4 mm 0.79" x 1.8" x 0.29"
Weight 85 g (3 oz)
Active detector area  10 x 10 mm 0.39" x 0.39"
Cable length  2.0 m (6.6 ft)
Connector  LEMO triax
Dose range  0.1 nGy–>100 kGy 12 nR–>11000 kR
Dose rate range  1 nGy/s–76 mGy/s 0.4 mR/h–31 kR/h
Dose per pulse  1 nGy/pulse–1000 Gy/pulse 12 nR/pulse–110 kR/pulse
Backscatter  None
 
 